| Название: | Comparison of theoretical fixation stability of three devices employed in medial opening wedge high tibial osteotomy: a finite element analysis |
| Авторы: | Golovakhа, М. L. Orljanski, W. Benedetto, K.-P. Panchenko, S. Büchler, P. Henle, P. Aghayev, E. Головаха, Максим Леонідович |
| Ключевые слова: | High tibial osteotomy Open wedge osteotomy Puddu plate TomoFix plate Tibial osteotomy fixation Comparative osteosynthesis stability |
| Дата публикации: | 2014 |
| Библиографическое описание: | Comparison of theoretical fixation stability of three devices employed in medial opening wedge high tibial osteotomy: a finite element analysis / М. L Golovakhа, W. Orljanski, K.-P. Benedetto, S. Panchenko, P. Büchler, P. Henle, E. Aghayev // BMC Musculoskeletal Disorders. - 2014. - Vol. 15. - Art. 230. - https://doi.org/10.1186/1471-2474-15-230. |
| Аннотация: | Background: Medial open wedge high tibial osteotomy is a well-established procedure for the treatment of
unicompartmental osteoarthritis and symptomatic varus malalignment. We hypothesized that different fixation
devices generate different fixation stability profiles for the various wedge sizes in a finite element (FE) analysis.
Methods: Four types of fixation were compared: 1) first and 2) second generation Puddu plates, and 3) TomoFix
plate with and 4) without bone graft. Cortical and cancellous bone was modelled and five different opening wedge
sizes were studied for each model. Outcome measures included: 1) stresses in bone, 2) relative displacement of the
proximal and distal tibial fragments, 3) stresses in the plates, 4) stresses on the upper and lower screw surfaces in
the screw channels.
Results: The highest load for all fixation types occurred in the plate axis. For the vast majority of the wedge sizes
and fixation types the shear stress (von Mises stress) was dominating in the bone independent of fixation type. The
relative displacements of the tibial fragments were low (in μm range). With an increasing wedge size this
displacement tended to increase for both Puddu plates and the TomoFix plate with bone graft. For the TomoFix
plate without bone graft a rather opposite trend was observed.
For all fixation types the occurring stresses at the screw-bone contact areas pulled at the screws and exceeded the
allowable threshold of 1.2 MPa for at least one screw surface. Of the six screw surfaces that were studied, the
TomoFix plate with bone graft showed a stress excess of one out of twelve and without bone graft, five out of
twelve. With the Puddu plates, an excess stress occurred in the majority of screw surfaces.
Conclusions: The different fixation devices generate different fixation stability profiles for different opening wedge
sizes. Based on the computational simulations, none of the studied osteosynthesis fixation types warranted an
intransigent full weight bearing per se. The highest fixation stability was observed for the TomoFix plates and the
lowest for the first generation Puddu plate. These findings were revealed in theoretical models and need to be
validated in controlled clinical settings. |