IRZSMU >
Кафедри >
Кафедра органічної і біоорганічної хімії >
Наукові праці. (Органічна хімія) >
Пожалуйста, используйте этот идентификатор, чтобы цитировать или ссылаться на этот ресурс:
http://dspace.zsmu.edu.ua/handle/123456789/13686
|
Название: | Method development and validation for the determination of residual solvents in quinabut API by using gas chromatography. Message 2 |
Авторы: | Golembiovska, O. Voskoboynik, O. Berest, G. Kovalenko, S. Logoyda, L. Берест, Галина Григорівна Коваленко, Сергій Іванович |
Ключевые слова: | Gas Chromatography Quinabut Residual Solvents Validation |
Дата публикации: | 2021 |
Библиографическое описание: | Method development and validation for the determination of residual solvents in quinabut API by using gas chromatography. Message 2 / O. Golembiovska, O. Voskoboynik, G. Berest, S. Kovalenko, L. Logoyda // Pharmacia. - 2021. - Vol. 68, № 1. - P53–59. - https://doi.org/10.3897/pharmacia.68.e52119 |
Аннотация: | The aim of study was to develop and validate a simple, precise and accurate method using gas chromatography for analysis of
residual solvents – acetone and 2-propanol – in quinabut API.
Materials and methods. All experiments were performed on a gas chromatographic system equipped with FID detector (Shimadzu
GC System) using the DB-624 (30 m × 0.32 mm ID, 3.0 μm film sickness) column as stationary phase. Nitrogen was used as carrier
gas with flow rate 7.5 mL/ min. Split ratio was 1:5, injector temperature was 140 °C, detector temperature was 250 °C, oven temperature was programmed from 40 °C (2 min) to 50 °C at 1 °C/min and then increased at a rate of 15 °C/min up to 215 °C; and maintained
for 2 min. All solutions were prepared using water as diluent.
Results. This proposed method is assessed for separation of residual solvent from quinabut with quantification. The obtained results
are compared with the corresponding specified limits of ICH standard guidelines. The method validation was done by evaluating
specificity, limit of detection (LOD) and limit of quantitation (LOQ), linearity, accuracy, repeatability, ruggedness, system suitability
and method precision of residual solvents as indicated in the ICH harmonized tripartite guideline. The separation between acetone
and 2-propanol peaks is 2.07. Hence method was found to be specific. The linear relationship evaluated across range of 15 to 180%
for acetone and 2-propanol of ICH specified limit of residual solvents. The graphs of theoretical concentration versus obtained concentration are linear and the regression coefficients ‘R’ for residual solvents were more than 0.9968. The values of LOD and LOQ were
much less than the lower limit of the concentration range and cannot affect the accuracy of the test. The technique was characterized
by high intra-laboratory accuracy at concentrations close to the nominal acetone and 2-propanol concentration. All solutions were
stable in water for at least 1 hour when stored at room temperature.
Conclusion. A simple, specific, accurate, precise and rugged gas chromatography method was developed and validated for the
quantification of residual solvents present in quinabut API through an understanding of the synthetic process, nature of solvents and
nature of stationary phases of columns. The residual solvents acetone and 2-propanol were determined. |
URI: | http://dspace.zsmu.edu.ua/handle/123456789/13686 |