Название: | Biomechanical Substantiation of the Strength and Stiffness of a Hip Joint Capsule Defect Fixation with Polypropylene Mesh |
Авторы: | Maslennikov, S. O. Panchenko, S. P. Golovakha, M. L. Масленніков, Сергій Олегович Панченко, С. П. Головаха, Максим Леонідович |
Ключевые слова: | endoprosthesis hip joint modeling polypropylene total hip arthroplasty |
Дата публикации: | 2020 |
Библиографическое описание: | Maslennikov S. O. Biomechanical Substantiation of the Strength and Stiffness of a Hip Joint Capsule Defect Fixation with Polypropylene Mesh / S. O. Maslennikov, S. P. Panchenko, M. L. Golovakha // Surgery Journal. - 2020. - Vol. 6, № 1. - С. Е11-е18. - https://doi.org/10.1055/s-0039-3402793 |
Аннотация: | Formulation of the problem Dislocation of the femoral component of the endoprosthesis is one of the most frequent complications of total hip replacement. The best
option for the “treatment” of dislocation of the hip endoprosthesis is to prevent the
development of primary instability. There are cases in which even with the correct
installation of the endoprosthesis components, dislocations arise due to the weakness
or defect of the capsular–ligament apparatus. Currently, many methods have been
developed to strengthen and restore the posterior structures of the capsule of the hip
joint with the help of auto- and allomaterials, which differ in both the fixation technique
and the characteristics of the materials themselves. In this paper, we propose a method
for restoring and strengthening the posterior structures of the capsule of the hip joint
using polypropylene-based graft implants. The purpose of this study is to, with the help
of specialized software, build a model of the capsule of the hip joint after capsulotomy
and to determine the stiffness capabilities of the defect covered with polypropylene
mesh.
Results The study was performed using a software package based on the finiteelement method. As a result of the performed calculations, pictures of the distribution
of the stress–strain state in the “head-capsule” system were obtained. To assess the
effectiveness of the method of closing the capsule, from the viewpoint of rigidity, as
the main characteristics, the values of the opening of the cut are selected.
Conclusions Under the kinematic loading of the model, the smallest values of the
opening of the section are obtained when it is closed by a grid. In the case of thread
fixation, the values were higher by 8.5%. However, the values of equivalent stresses,
both in the capsule and in the head, in the model with the grid turned out to be the
largest. These stresses were higher by 23.8% in the capsule and by 60.4% in the head
than the same values for the thread fixation model. The obtained results indicate that
the model with a grid is more rigid in the considered fixation variants. |